DC bias current source MICROTEST 6243 (40A)

Price: Contact

Model 6243
Manufacturer MICROTEST
Power supply 88-264Vac
Weight 20 Kg
Size 356 x 145 x 644 mm
Condition Stocking
Warranty 12 months

DC bias current source MICROTEST 6243 (40A) from manufacturer MICROTEST provides a single output current of 40A.

I. Specification of DC bias source MICROTEST 6243 (40A)

DC bias current source MICROTEST 6243 (40A) is designed for performing magnetic saturation current (Isat) and temperature rise current analysis (Irms) on inductors.

To conduct such analysis, you have the option to purchase the system along with the 6632 Impedance Analyzer or the 6355/6356 LCR Meter.

  • If you choose the 6632 Impedance Analyzer, it has a frequency range of 100Hz-10MHz and can handle a maximum output current of 320A using 8 units.
  • For the 6355 LCR Meter, it has a frequency range of 100Hz-200kHz and can handle a maximum output current of 320A using 8 units.
  • The 6356 LCR Meter has a frequency range of 100Hz-500kHz and can handle a maximum output current of 320A using 8 units.

Nguon-hien-tai-thien-vi-dc-microtest-6243-40a_01

Main features of MICROTEST 6243:

  • Current and frequency graphic scanning analysis
  • Temperature-rising scan function can solve the problems of overheating a DUT to burn
  • DCR Measurement function
  • Long-term consecutive maximum power output
  • Interchangeable bi-direction current function
  • Frequency response
  • 100Hz-10MHz (With DC Bias Current 6223/6243), 100Hz-30MHz (With DC Bias Current 6225)
  • DC Bias Current Max.320A (6243)
  • Direct Handler interfaces control through LCR power meter.

Applications:

  • Components: High current power inductor, common mode choke, mini molding choke, high power components of EV charging connector
  • Electric Vehicles: Electric supercharger system.

II. Key features of DC bias current source MICROTEST 6243 (40A)

How DC bias current helps inductor testing?

  • Ferromagnetic materials (like iron) are composed of microscopic regions called magnetic domains. The stronger the external magnetic field H, the more the domains align, yielding a higher magnetic flux density B. Eventually, The magnetization remains nearly constant, and is said to have saturated. The domain structure at saturation depends on the temperature.

Dc-bias-current-source-microtest-6243-40a

What is magnetic saturation?

  • In electronic circuits, transformers and inductors with ferromagnetic cores operate nonlinearly when the current through them is large enough to drive their core materials into saturation. This means that their inductance and other properties vary with changes in drive current. In linear circuits this is usually considered an unwanted departure from ideal behavior.
  • Magnetic saturation current is called Isat, and the temperature rise current is called I rms. When the transformer and the inductor pass a large current in the actual circuit operation, the magnetic field of the magnetic core will produce magnetic saturation, which will cause the inductance characteristic to decline. Therefore, the R&D engineer will set the current value of the inductance reduction allowable range.

Dc-bias-current-source-microtest-6243-40a

Option 6632+6243 DC bias curretnt test system:

  • Supports magnetic saturation current analysis measurement function.

Dc-bias-current-source-microtest-6243-40a

Isat (Magnetic saturation current curve):

Dc-bias-current-source-microtest-6243-40a

Irms (Rated current curve):

Dc-bias-current-source-microtest-6243-40a

Control inductance:

  • Using a DC Bias current source to apply a 200A bias current to the inductor, the inductance decreased from 2.06983uH to 1.02845uH.

Dc-bias-current-source-microtest-6243-40a

Equivalent circuit analysis (Option):

  • It has seven different models, combine with different types of parameters ( R, L, C) help customers model the impedance versus frequency characteristics and self-resonant frequency (SRF).

Dc-bias-current-source-microtest-6243-40a

Magnetic Material Test Fixture (FX-0000C8):

  • Using the magnetic material test fixture for measuring of permeability of various toroidal cores or ferrite cores and electromagnetic shielding coating materials, 6632 built-in formula to directly calculate the permeability coefficient value μr.
Dc-bias-current-source-microtest-6243-40a Dc-bias-current-source-microtest-6243-40a

 

List mode:

  • 50 groups of list mode setting (Each group contains up to 15 steps).

Dc-bias-current-source-microtest-6243-40a

III. System selection MICROTEST 6243 DC bias source

DC Bias Current Source

Model

6243

DC Bias Current Max.

320A

units of the DC Bias

8

Power consumption

640W Max.

Output Current (unit)

40A

Option LCR Meter/Impedance Analyzer

6632/6356/6355

Option LCR Meter/Impedance Analyzer

Frequency Response

6632

100Hz-10MHz

6356

100Hz-500kHz

6355

100Hz-200kHz

Technical specifications of DC bias current source MICROTEST 6243 (40A)

DC Bias Current Source

6243

 Accuracy (6632)

0.000A-1.000A 1%+5mA

1.001A-5.000A 2%

5.001A-20.000A 3%

DCR Test

Constant power output

Current switch

 

System:

   Power Supply

Voltage:88-264Vac

Frequency:47-63Hz

  Environment

Temperature:10℃-40℃

Humidity:20-90%RH

Dimension

356x145x644 mm (W*H*D)/6243

  Weight

20 Kg (6243)

Display

Interface

RS-232、Handler

Review DC bias current source MICROTEST 6243 (40A)

5 0% | 0 review
4 0% | 0 review
3 0% | 0 review
2 0% | 0 review
1 0% | 0 review
Review DC bias current source MICROTEST 6243 (40A)
Choose a image
0 character ( Minimum of 10)
    +

    There are no reviews yet.

    No comments yet

    SIGN UP FOR A CONSULTATION

    Please fill in the form below and send your comments and questions to GMGA, we will respond to your email as soon as possible.

      COMMIT

      100% new product

      COMMIT

      100% refund if the product is of bad quality and not as promised

      24/7 SUPPORT

      Hotline: 0845 969 336

      DELIVERY

      Nationwide