Hipot tester MICROTEST 7630 (2 Channel)
Price: Contact
Model | 7630 |
Manufacturer | MICROTEST |
Power supply | 90Vac-264Vac |
Weight | 14 Kg |
Size | 365 x 145 x 430 mm |
Condition | Stocking |
Warranty | 12 months |
Hipot tester MICROTEST 7630 (2 Channel) has a maximum AC withstand voltage of 5000V with a leakage current of 30mA, a maximum DC withstand voltage of 6000V with a leakage current of 10mA, and a maximum insulation resistance of 12000MΩ.
I. Specification of Safety Tester MICROTEST 7630 (2 Channel)
- Precision Hipot tester MICROTEST 7630 (2 Channel) is a 3-in-1 test instrument that supports AC withstand voltage, DC withstand voltage, and insulation resistance testing functions.
- Hipot meter MICROTEST 7630 supports PC connection software for remote programming and storage of test data. It can also be linked to a barcode scanner for barcode scanning, facilitating production traceability management.
Features of Hipot tester MICROTEST 7630:
- 2 Channels
- Hi pot ACW 5KV/30mA
- Hi pot DCW 6KV/10mA
- Insulation 1KV/12000MΩ
- Arcing detection
- Measuring Time (0.1 ~ 999s)
- Adjustable ramp time (0.1 ~ 10s)
- Safety Switch(INTER LOCK)
- USB Host storages setting files and can update firmware
- Support Remote control signal.
Applications:
- Transformers, motors, capacitors, inductors, wires and wave filter.
II. Main features Hipot tester MICROTEST 7630 (2 Channel)
MICROTEST 7630 (2 Channel) HI-POT tester safety mechanism:
- Output safety switch.
- Switch on the inter lock before testing.
Instrument will stop testing when open loop happened.
Synchronization display INTER LOCK.
Safety certification testing items:
- AC Hi-pot
- DC Hi-pot
- Insulation resistance
- Current leakage.
AC/DC Withstanding voltage test:
- Difference between AC and DC
AC Hipot Test |
Most DUT have parasitic capacitance. With AC test, might not able to charge the parasitic capacitance. Therefore, this will create constantly current go through. |
DC Hipot Test |
The parasitic capacitance will be charged under DC hi-pot test. Therefore, the current will reduce to almost 0. |
- Advantages for AC Test
AC Hipot Test |
AC test is same as our normal usage electricity. Also, AC is able to do positive and negative polarity test.
AC test won’t charge the parasitic capacitance on the DUT, so it won’t create instantaneous peak current. Do not need to discharge the DUT after testing. |
DC Hipot Test |
The parasitic capacitance on DUT has been charged under DC test, so the rest leakage current is the real current from DUT. |
- Disadvantage of AC/DC Hi-pot test
AC Hipot Test |
If the DUT has huge parasitic capacitance, the current we detect will bigger then the actual current. |
DC Hipot Test |
Testing Voltage has to increase from 0. |
Hi-pot test time:
- The testing voltage is refer to the condition of DUT while doing Hi-pot test. If the voltage setting is too low, the insulation material defect might not be detected. If the voltage setting is too high and the testing time is too long, it may cause the permanent damage for the material.
- Normally setting for the safety standard.
Formula →Testing Voltage= DUT recommend voltage × 2+1000V
EX : DUT standard :240V →Setting test voltage:1480V.
Common way to safe time for testing:
- Testing time for most products is 60s. If there are more than one test on one item, it will cost a lot of time. This will lower the efficiency og the product line.
- Therefore, the product line usually shorter the testing time to 1~2 sec. and higher 10%~20% testing voltage.
Insulation resistance & Arc detection:
What is Insulation resistance test?
- VDE and TUV safety standard request insulation resistance test for some specific product before hi-pot test. The reason is to make sure the insulation will not be damaged while doing hi-pot testing. Insulation resistance test is the key item to evaluate the material.
- Hi-pot test is judge by leakage current, but IR test is judge by resistance, and usually it’s higher than 1 MΩ. It tests the DUT’s resistance between both terminal.
Why do we need to do IR test?
- Confirm the performance of the structure of the insulation.
- The insulation structure should have higher insulation resistance rate by using better material.
- For the insulation material, the protection against electric shock of the switch must be guaranteed.
- Confirm the insulation structure can pass the reliability test(heat durability, insulation)
- The importance of insulation resistance
- Before assemble the product, IQC has to do the IR test for every components, such as switches, transformers, resistors, capacitors, inductors, PCB, wire or cable to make sure the quality.
Leakage current test?
→Measuring Tiny AC Leakage Current
- DUT has tiny parasitic capacitance. This characteristic will create leakage current by resistance. Capacitor is a resistance component under AC voltage. This may cause the test result include the leakage current inside the instrument. If the DUT has low leakage current, we must consider capacitance , frequency, and the voltage we put.
Arc detection?
- The surface or interior of the insulation material might discharge under high voltage test. Because of the lost of insulation performance, DUT may have temporary of discontinuity discharge.
- Because that temporary discharge can’t be judge by leakage current, so Hi-pot tester will judge the defect product by the changing rate of testing voltage or leakage current.
- Arc test is a geometric not a constant test. Also, the detection will be affect by the wire’s impedance or capacitance between the detectors. The rate of change of the current will be affect in the Arc detection.
Technical specifications of Hipot tester MICROTEST 7630 (2 Channel)
Model |
7630 |
Test Channel |
2 |
AC Output Voltage |
5000V |
DC Output Voltage |
6000V |
Insulation Resistance |
12000MΩ |
AC Hipot |
|
Output Voltage |
10V-5000V |
Voltage Resolution |
1V |
Voltage Accuracy |
±(3% of setting+5V) |
Rated Current (Max.) |
30mA |
Upper Leakage Current |
0.001-31mA |
Leakage Current Resolution |
0.001mA |
Leakage Current Accuracy |
±(3% of reading +10uA) |
Arcing Detection |
Detect gears from 0-20 (1 is the most sensitive) |
DC Hipot |
|
Output Voltage |
10V-6000V |
Voltage Resolution |
1V |
Voltage Accuracy |
±(3% of setting+5V) |
Rated Current (Max.) |
10mA |
Upper Leakage Current |
0.001-11mA |
Leakage Current Resolution |
±(3% of reading +10uA) |
Leakage Current Accuracy |
0.001mA |
Arcing Detection |
Detect gears from 0-10 (1 is the most sensitive) |
Insulation Resistance |
|
Output Voltage |
10V-1000Vdc |
Voltage Resolution |
1V |
Voltage Accuracy |
±(3% of setting+5V) |
Insulation Resistance Range |
1-12000MΩ |
IR Resolution |
1MΩ |
IR Accuracy |
100-500V, 1-1000MΩ (±5% of reading+1MΩ) |
Measuring Time |
0.1-999s |
Ramp Time |
0.1-10s |
Measuring Time Resolution |
0.1s |
System:
PLC Remote Control |
Test、About |
PLC Output Signal |
PASS、FAIL、HV Output、Testing |
Safety Switch |
When testing, you need to short-circuit the INTER LOCK on the rear of the instrument to output the test voltage |
List Mode |
Each group contains up to 16 steps |
Built-in Storage |
30 sets setting files |
USB Host Storage |
Storage setting files and update firmware |
Power Supply |
Voltage:90Vac-264Vac Frequency:50/60Hz ±5% |
Power consumption |
300VA |
Environment |
Temperature:5℃-40℃ Humidity:0-80%RH |
Dimension |
365x145x430 mm (W*H*D) |
Weight |
14 Kg |
Interface |
RS-232、Remote、USB Host、USB Device、EXT. I/O |
Display |
4.3″ TFT LCD (480*272 ) |
No comments yet
SIGN UP FOR A CONSULTATION
Please fill in the form below and send your comments and questions to GMGA, we will respond to your email as soon as possible.
COMMIT
100% new product
COMMIT
100% refund if the product is of bad quality and not as promised
24/7 SUPPORT
Hotline: 0845 969 336
DELIVERY
Nationwide
Review Hipot tester MICROTEST 7630 (2 Channel)