Labculture ® Biological Safety Cabinets Class II Type A2 BSC
Price:
Model | |
Manufacturer | |
Power supply | |
Weight | |
Size | |
Condition | |
Warranty |
I. Specification microbiological safety cabinet Labculture ® Class II Type A2 BSC
- Esco Labculture® Class II Biological Safety Cabinets combine the latest generation of energy-efficient DC ECM motor technology with Esco's more than 25 years of biosafety cabinet design experience.
- Labculture ® Class II Type A2 BSC combines multiple design, construction and proprietary considerations to deliver optimum operator efficiency, product and environmental protection.
- Labculture® is equipped with 2 ULPA (Ultra Low Penetrating Air) filters that operate at a typical efficiency of 99.999%, surpassing conventional HEPA (High Efficiency Air) filters in pollution protection. . Esco Labculture® features a 10 degree inclined front panel for maximum operator comfort. One-piece stainless steel work area and no screws or connectors in the work area facilitate easy cleaning after the test.
Sentinel™ Gold Microprocessor Control System:
- There are audible and visual alarms
- Display all safety information on one screen
- Centered and tilted down for easy access and viewing
- Selectable Quick Start mode for quick operation.

Efficient dual DC-ECM blower system:
- Powered by the latest generation, more efficient DC ECM motor legacy ECM and VFD engines
- 70% energy saving compared to AC . motor
- Stable airflow, despite voltage fluctuations and filter loads
- Night setting mode to further reduce power consumption by 60%.


Filter ULPA:
- 10 x filtration efficiency compared to HEPA filter
- Create an ISO Class 3 workspace instead of the industry standard ISO Class 5
- 99.999% at 0.1 to 0.3 micron, ULPA according to IEST-RP-CC001.3 USA or >99.999% at MPPS, H14 according to EN 1822 EU.
- Labculture ® Biological Safety Cabinets Class II Class A2 BSC use ULPA (according to IEST-RP-CC001.3) / H14 filter on EN 1822 instead of HEPA H13 filter used on many BSCs on the market.


Cabinet structure:
- Stainless steel side walls with side capture zones and negative pressure for optimized containment
- Easy access to service devices
- One-piece work tray, spill resistant and easy to clean.

Angled Drain Pan:
- Easy to clean
- Contains no contaminant.

Adjustable UV timer:
- Easily adjust to desired minutes or hours
- Extend the using time of the UV lamp, if not turn ON the lamp overnight.

ISOCIDE™ Coating:
- Silver ion impregnated powder coat
- Inhibits microbial growth for improved safety.

Cabinet filter system LA2 and LR2 CLASS II TYPE A2:
- Ambient air is pulled through the front grille to create an inlet without entering the work surface. The inlet is joined by a downward half, to create a front air curtain that is fine-tuned to create a large performance envelope. The combined airflow passes through the rear air column towards the blower.
- About ⅓ of the air in the common room is expelled through the ULPA filter into the room. ⅔ remaining air is passed through the downflow ULPA filter and into the work area as vertical laminar flow air to create an ISO Class 3 work surface and prevent cross contamination.
- Near the work surface, the downflow separates. About half belongs to the front grille and half belongs to the rear grille. A small portion goes into the side snap areas to prevent dead air angles (small blue arrows)
- Optimized design for great performance, providing operator and product protection at wide Inflow and Downflow variations from Nominal point.
Isocide ™ Antibacterial Powder Coating:
- Removes 99.9% of bacteria on surfaces within 24 hours of contact.

II. Advantages of Labculture ® Biological Safety Cabinets Class II Type A2 BSC
- The RS 232 data output port allows remote monitoring of the cabinet's operating parameters.
- The Sentinel TM Gold microprocessor controller is designed to be centered and angled, displaying all safety information on one screen.
- Easy-to-clean one-piece wall with side power outlet and staggered service fixtures
- Recessed one-piece work tray easily spills
- Ergonomically designed raised armrests help block the grille and create a comfortable working posture
- The angled drain pan is easy to clean and contains no contaminants.
- Airflow sensor monitors airflow in real time for safety, alerting users if airflow is insufficient
- Energy efficient DC ECM motor saves 70% energy compared to AC Motor.
- Equipped with ULPA filter providing better filtration than HEPA filter with efficiency > 99.999% conventional at 0.1 micron.
- Dynamic Chamber TM (the void of the blower and the side walls surrounded by negative pressure) prevents contaminants from escaping to the outside.
- Isocide TM powder coating inhibits microbial growth on external surfaces, prevents contamination and improves safety.
III. Specifications Labculture ® Biological Safety Cabinets Class II Type A2 BSC
TECHNICAL SPECIFICATIONS |
||||||
Labculture® Class II A2 |
LA2-3A_-E | LA2-4A_-E | LA2-5A_-E | LA2-6A_-E | LA2-8A_-E | |
Labculture® Reliant Class II A2 | LR2-3S_-E | LR2-4S_-E | LR2-5S_-E | LR2-6S_-E |
LR2-8S_-E |
|
Nominal Size |
0.9 meter (3') | 1.2 meter (4') | 1.5 meter (5') | 1.8 meter (6') | 2.4 meters (8') | |
External Dimensions * (W x D x H) | 1115 x 852 x 1540 mm (44.0" x 33.5" x 60.6") | 1420 x 852 x 1540 mm (56.0" x 33.5" x 60.6") | 1725 x 852 x 1540 mm (68.0" x 33.5" x 60.6") | 2030 x 852 x 1540 mm (80.0" x 33.5" x 60.6") |
2600 x 852 x 1540 mm (102.4" x 33.5" x 60.6") |
|
Internal Dimensions (W x D x H) |
970 x 623 x 670 mm (38.2" x 24.5" x 26.4") | 1270 x 623 x 670 mm (50.0" x 24.5" x 26.4") | 1570 x 623 x 670 mm (61.8" x 24.5" x 26.4") | 1870 x 623 x 670 mm (73.6" x 24.5" x 26.4") | 2440 x 623 x 670 mm (96.0" x 24.5" x 26.4") | |
Usable Work Area | 0.45 m2 (4.8 sq.ft.) | 0.6 m2 (6.5 sq.ft.) | 0.75 m2 (8.1 sq.ft.) | 0.9 m2 (9.7 sq.ft.) |
1.2 m2 (13 sq.ft.) |
|
Tested Opening |
229 mm (9") | 229 mm (9") | 229 mm (9") | 203 mm (8") | 203 mm (8") | |
Working Opening | 274 mm (10.8") | 274 mm (10.8") | 274 mm (10.8") | 248 mm (9.8") |
248 mm (9.8") |
|
Average Airflow Velocity |
Inflow | 0.53 m/s (105 fpm) | ||||
Downflow | 0.35 m/s (70 fpm) | 0.35 m/s (70 fpm) | 0.35 m/s (70 fpm) | 0.33 m/s (65 fpm) |
0.33 m/s (65 fpm) |
|
Airflow Volume |
Inflow | 424 m3 /h (251 cfm) | 555 m3 /h (328 cfm) | 686 m3 /h (406 cfm) | 724 m3 /h (426 cfm) | 945 m3 /h (560 cfm) |
Downflow | 628 m3 /h (363 cfm) | 822 m3 /h (476 cfm) | 1016 m3 /h (588 cfm) | 1210 m3 /h (700 cfm) |
1579 m3 /h (914 cfm) |
|
Exhaust |
424 m3 /h (251 cfm) | 555 m3 /h (328 cfm) | 686 m3 /h (406 cfm) | 724 m3 /h (426 cfm) | 945 m3 /h (560 cfm) | |
Required Exhaust with Optional Thimble Exhaust Collar | 529 m3 /h (311 cfm) | 764 m3 /h (450 cfm) | 1116 m3 /h (657 cfm) | 1164 m3 /h (685 cfm) |
1540 m3 /h (913 cfm) |
|
Static Pressure for Optional Thimble Exhaust Collar |
32 Pa / 0.12 in H2O | 49 Pa / 0.19 in H2O | 62 Pa / 0.24 in H2O | 79 Pa / 0.31 in H2O | 100 Pa / 0.40 in H2O | |
ULPA Filter Typical Efficiency |
>1000lux (>93 foot-candles) |
|||||
Cabinet Construction |
Main Body | Electro-galvanized steel with white oven-baked epoxy-polyester Isocide™ antimicrobial powder-coated finish, 1.5 mm (0.06") / 16 gauge thick | ||||
Work Zone |
Stainless steel Type 304 with No.4 finish, 1.5 mm (0.06") / 16 gauge thick |
|||||
Electrical |
Full Load Amps 230 V | 10 A | 10 A and 5 A | |||
Full Load Amps 115 V | 13 A |
13 A and 8 A |
||||
Heat Load |
853 BTU/Hr | 972 BTU/Hr | 1177 BTU/Hr | 1297 BTU/Hr | 1774 BTU/Hr | |
Nominal Power Consumption | 250 W | 285 W | 345 W | 380 W |
520 W |
|
Net Weight*** |
243 Kg (536 lbs) | 283 Kg (624 lbs) | 350 Kg (772 lbs) | 426 Kg (939 lbs) | 580 Kg (1279 lbs) | |
Shipping Weight*** | 292 Kg (644 lbs) | 345 Kg (761 lbs) | 410 Kg (904 lbs) | 486 Kg (1072 lbs) |
640 Kg (1411 lbs) |
|
Shipping Dimensions, Maximum (W x D x H)*** |
1200 x 950 x 1900 mm (47.2" x 37.4" x 74.8") | 1550 x 950 x 1900 mm (61.0" x 37.4" x 74.8") | 1950 x 950 x 1900 mm (76.8" x 37.4" x 74.8") | 2150 x 950 x 1900 mm (84.6" x 37.4" x 74.8") | 2720 x 950 x 1900mm (84.6" x 37.4" x 74.8") | |
Shipping Volume, Maximum*** | 2.17 m³ (77 cu.ft.) | 2.80 m³ (99 cu.ft.) | 3.52 m³ (124 cu.ft.) | 3.88 m³ (137 cu.ft.) |
4.91 m³ (173 cu.ft.) |
Model LA2 and LR2 Biological Safety Cabinet Engineering Drawing:

Optional Exhaust Collar Positions for Thimble-Ducting for LA2 and LR2 Models:
V. Accessories for Labculture ® Biological Safety Cabinets Class II LA2, LB2 and LR2
Accessories for LA2, LB2 and LR2 Biological Safety Cabinets |
|||||||
Cabinet |
Stainless Steel Side Wall | LA2-3A1-E 2010705 | LA2-4A1-E 2010670 | LA2-5A1-E 2010671 | LA2-6A1-E 2010672 | LA2-8A1-E 2010928 | |
LA2-3A2-E 2010706 | LA2-4A2-E 2010691 | LA2-5A2-E 2010692 | LA2-6A2-E 2010693 |
LA2-8A2-E 2011205 |
|||
LA2-3A3-E 2010707 |
LA2-4A3-E 2010685 | LA2-5A3-E 2010686 | LA2-6A3-E 2010687 | LA2-8A3-E 2010921 | |||
LB2-3B1-E 2010708 | LB2-4B1-E 2010673 | LB2-5B1-E 2010674 | LB2-6B1-E 2010682 | ||||
LB2-3B2-E 2010709 |
LB2-4B2-E 2010694 | LB2-5B2-E 2010695 | LB2-6B2-E 2010696 | LB2-8B2-E 2011206 | |||
LB2-3B3-E 2010710 | LB2-4B3-E 2010688 | LB2-5B3-E 2010689 | LB2-6B3-E 2010690 | ||||
LR2-4S1-E 2010845 | LR2-5S1-E 2010847 | LR2-6S1-E 2010849 | |||||
LR2-3S2-E 2010701 | LR2-4S2-E 2010702 | LR2-5S2-E 2010703 | LR2-6S2-E 2010704 |
LR2-8S2-E 2011006 |
|||
LR2-3S3-E 2010913 |
LR2-4S3-E 2010851 | LR2-5S3-E 2010853 | LR2-6S3-E 2010855 | ||||
Exhaust Ducting | Anti-blowback Valve 10 inches (LA2 & LR2 only | EG Powder Coated |
ABBV-10P 5170352 |
||||
304 Stainless Stee |
ABBV-10S 5170354 | ||||||
Anti-blowback Valve 12 inches (LB2 only) | EG Powder Coated |
ABBV-12P 5170353 |
|||||
304 Stainless Steel |
ABBV-12S 5170355 | ||||||
Exhaust Damper |
B2-DAMPER 5170104 |
||||||
Exhaust Collar (LA2 & LR2 only) |
ECO-LA23-MK3-LH 5170097 | ECO-LA24-MK3-LH 5170099 | ECO-LA25-MK3-LH 5170101 | ECO-LA26-MK3-LH 5170102 | ECO-LA28-MK3-LH 5170536 | ||
Inlet Collar (LB2 only) | ICO-LB23 5170320 | ICO-LB24 5170263 | ICO-LB25 5170316 | ICO-LB26 5170322 |
ICO-LB28 5170692 |
||
Pre-filter (LB2 only |
PF-2 6090001 | ||||||
Electrical Outlet | UV Lamp | UV-15A-L 5170251 | UV-30A-L 5170255 |
UV-15A-L (x2) 5170251 |
|||
IV Bar |
IV-955 5170276 | IV-1260 5170277 | IV-1265 5170278 | IV-1870 5170279 | |||
Direct Mounted / GFCI |
EO-_ |
||||||
Service Fixtures |
EU SF-Gas-40 mm |
SF-1G40 5170002 | |||||
EU SF-Vacuum-40 mm |
SF-1V40 5170003 |
||||||
EU SF-Air-40 mm |
SF-1A40 5170006 |
||||||
EU SF-Nitrogen-40 mm |
SF-1N40 51700011 | ||||||
EU SF-Water-40 mm |
SF-1W40 51700017 |
||||||
EU SF-Universal-40 mm |
SF-2U40 51700018 | ||||||
Support Stands, Ships Flat | Fixed Stand with Leveling Feet, 28” height | SPL-3A0 Gen 2 5130188 | SPL-4A0 Gen 2 5130189 | SPL-5A0 Gen 2 5130190 | SPL-6A0 Gen 2 5130163 |
SPBL-8A0 5131286 |
|
Fixed Stand with Leveling Feet, 34” height |
SPL-3B0 Gen 2 5131092 | SPL-4B0 Gen 2 5130151 | SPL-5B0 Gen 2 5131093 | SPL-6B0 Gen 2 5131094 | SPBL-8B0 5131287 | ||
Fixed Stand with Caster Wheels, 28” height | SPC-3A0 Gen 2 5130155 | SPC-4A0 Gen 2 5130152 | SPC-5A0 Gen 2 5130162 | SPC-6A0 Gen 2 5130154 |
SPC-8A0 Gen 2 5131122 |
||
Fixed Stand with Caster Wheels, 34” height |
SPC-3B0 Gen 2 5130165 | SPC-4B0 Gen 2 5130166 | SPC-5B0 Gen 2 5130167 | SPC-6B0 Gen 2 5130168 | SPC-8B0 Gen 2 5131123 | ||
Telescopic Stand with Leveling Feet, 1” adjustment | STL-3A0 5130050 | STL-4A0 5130051 | STL-5A0 5130052 | STL-6A0 5130053 |
STL-8A0 5130054 |
||
Telescopic Stand with Caster Wheels, 1” adjustment |
STC-3A0 5130055 | STC-4A0 5130056 | STC-5A0 5130057 | STC-6A0 5130058 | STC-8A0 5130059 | ||
Motorized Height Stand with Caster Wheels, 39.5” height | SPM-3A_ | SPM-4A_ | SPM-5A_ | SPM-6A_ |
SPM-8A_ |
||
Misc |
Arm Rest Padding | MEWREST 5170127 | |||||
Foot Rest |
FT-REST 5170492 |
||||||
Laboratory Chair |
ME-LD-AR360 1150006 | ||||||
IQ OQ Protocol |
9010179 |
V. Video
***See more videos of Class I, Class II, Class III biological safety cabinets at: Gmga Measuring Youtube.No comments yet
SIGN UP FOR A CONSULTATION
Please fill in the form below and send your comments and questions to GMGA, we will respond to your email as soon as possible.

COMMIT
100% new product

COMMIT
100% refund if the product is of bad quality and not as promised

DELIVERY
Nationwide
Review Labculture ® Biological Safety Cabinets Class II Type A2 BSC
There are no reviews yet.