Microbiological safety cabinet Esco Airstream® Class II Type A2 (AC2-NS) Certified to NSF-49
Price:
Model | |
Manufacturer | |
Power supply | |
Weight | |
Size | |
Condition | |
Warranty |
I. Specification Esco Airstream® Class II Class A2 Biological Safety Cabinets (AC2-NS)
Sentinel™ Gold Microprocessor Control System:
- There are audible and visual alarms
- Display all safety information on one screen
- Centered and tilted down for easy access and viewing
- Selectable Quick Start mode for quick operation.
Energy-saving motor DC-ECM blower:
- Esco Airstream® Gen 3 (AC2) Class II microbiological safety cabinet the most energy efficient BSC - D Series in the world, 70% energy saving compared to AC motors
- Stable airflow, despite voltage fluctuations and Loading filter
- Night setting mode to further reduce power consumption by 60%.
Filter ULPA:
- 10 times filtration efficiency compared to HEPA filter
- 99.999% at 0.1 to 0.3 micron, ULPA according to IEST-RP-CC001.3 USA or > 99.999% at MPPS, H14 according to EN 1822 EU.
One-piece wall:
- Easy access to service equipment and stores
- Large radius for easy cleaning.
One-piece work tray:
- Tunnel to prevent overflow
- Easy-to-clean sloped perimeter.
Raise the armrest:
- Helps prevent grilles
- Comfortable working posture.
Isocide ™ Antibacterial Powder Coating:
- Removes 99.9% of bacteria on surfaces within 24 hours of contact.
RS 232 serial interface port and voltage-free contact:
- Send operational information to the Building Management System (BMS)
- Zero Volt Relay contact for ON/OFF exhaust blower and building warning signal.
Powder coating outside:
- Silver ion impregnated powder coat
- Inhibits microbial growth for improved safety.
Removable paper tray:
- Easy to clean
- Optional pre-filter can be installed.
Cabinet filter system:
- Ambient air is drawn through the front grille to create an inlet without passing over the work surface. The inlet is joined by a downward half, to create a front air curtain that is fine-tuned to create a large performance envelope. The combined airflow passes through the rear air column towards the blower.
- About ⅓ of the air in the common room is expelled through the ULPA filter into the room. ⅔ remaining air is passed through a downflow ULPA filter and into the work area as a vertically layered air stream to create an ISO Class 3 work surface and prevent cross contamination.
- Near the work surface, the downflow separates. About Half is for the front grille, and half is for the rear grille. A small portion goes into the side snap areas to prevent dead air angles (small blue arrows).
- Optimized design for great performance, providing operator and product protection at wide Inflow and Downflow variations from Nominal point.
Standards compliance:
- Biological safety cabinets: NSF / ANSI 49 NSF*
- Air quality: ISO 14644.1, Class 3, Worldwide JIS B9920, Class 3, Japan JIS BS5295, Class 3, Japan US Fed Std 209E, Class 1 USA.
- Filter: EN-1822 (H14), Europe IEST-RP-CC001.3, USA IEST-RP-CC007, USA IEST-RP-CC034.1, USA.
- Electrical safety: UL-61010A-1, USA CSA22.2, No.1010-192, Canada EN-61010-1, Europe IEC61010-1, Worldwide.
II. Specifications microbiological safety cabinet Esco Airstream® Class II Type A2 (AC2-NS)
TECHNICAL SPECIFICATIONS |
|||||
Model |
220-240 VAC, 50/60Hz | AC2-3S8-NS 2010946 | AC2-4S8-NS 2010747 | AC2-5S8-NS 2010978 | AC2-6S8-NS 2010963 |
110-130 VAC, 50/60 Hz | AC2-3S9-NS 2010945 | AC2-4S9-NS 2010752 | AC2-5S9-NS 2010977 |
AC2-6S9-NS 2010925 |
|
External Dimensions (W x D x H) |
1035 x 753 x 1400 mm (40.7” x 29.6” x 55.1”) | 1340 x 753 x 1400 mm (52.8” x 29.6” x 55.1”) | 1645 x 753 x 1400 mm (64.8” x 29.6” x 55.1”) | 1950 x 753 x 1400 mm (76.8” x 29.6” x 55.1”) | |
Gross Internal Dimensions (W x D x H) | 915 x 580 x 660 mm (36.0” x 22.8” x 26.0”) | 1220 x 580 x 660 mm (48.0” x 22.8” x 26.0”) | 1525 x 580 x 660 mm (60.0” x 22.8” x 26.0”) |
1830 x 580 x 660 mm (72.0” x 22.8” x 26.0”) |
|
Usable Work Area |
0.42 m2 (4.5 ft2) | 0.56 m2 (6.0 ft2 ) | 0.70 m2 (7.5 ft2 ) | 0.86 m2 (9.3 ft2) | |
Tested Opening |
203 mm (8”) |
||||
Average Inflow Velocity |
0.53 m/s (105 fpm) | ||||
Average Downflow Velocity |
0.30 m/s (60 fpm) |
||||
Airflow Volume |
Inflow | 354 cmh (208 cfm) | 473 cmh (280 cfm) | 591 cmh (348 cfm) | 709 cmh (417 cfm) |
Downflow | 553 cmh (325 cfm) | 738 cmh (434 cfm) | 922 cmh (543 cfm) |
1107 cmh (652 cfm) |
|
Exhaust |
354 cmh (208 cfm) | 473 cmh (278 cfm) | 591 cmh (348 cfm) | 709 cmh (417 cfm) | |
Required Exhaust with Optional Thimble Exhaust Collar | 531 cmh (313 cfm) | 710 cmh (418 cfm) | 887 cmh (522 cfm) |
1064 cmh (626 cfm) |
|
Static Pressure For Optional Thimble Exhaust Collar |
32 Pa / 0.12 in H2O | 45 Pa / 0.18 in H2O | 57 Pa / 0.23 in H2O | 68 Pa / 0.27 in H2O | |
ULPA Filter Typical Efficiency |
>99.999% at 0.1 to 0.3 micron, ULPA as per IEST-RP-CC001.3 USA >99.999% at MPPS, H14 as per EN 1822 EU >99.99% on 0.3 micron size particles by PAO test (filter leakage test according to NSF49) |
||||
Sound Emission per NSF / ANSI 49* |
57.5 dBA | 58.5 dBA | 60.5 dBA | ||
Fluorescent Lamp Intensity | 982 lux (91 foot candles) | 1079 lux (100 foot candles) | 1040 lux (97 foot candles) |
1035 lux (96 foot candles) |
|
Cabinet Construction |
Main body | 1.2 mm (0.05") / 18 gauge EG Steel with IsocideTM Oven-Baked Epoxy-Polyester Powder Coating | |||
Work Zone |
1.5 mm (0.06”) / 16 gauge, SS 304, 4B finish |
||||
Side Walls |
1.5 mm (0.06”) / 16 gauge, SS 304, 4B finish | ||||
Electrical 220-240 VAC, 50/60 Hz | Cabinet Full Load Amps (FLA) | 8.5 | 8.6 | 9 |
12 |
Heat Load (BTU / Hr |
472 | 569 | 613 | 849 | |
Nominal Power Consumption (W) | 150 | 181 | 195 |
270 |
|
Electrical 110-130 VAC, 50/60 Hz |
Cabinet Full Load Amps (FLA) | 11 | 11.5 | 12.5 | 15 |
Heat Load (BTU / Hr) | 503 | 628 | 698 |
999 |
|
Nominal Power Consumption (W) |
160 | 200 | 222 | 318 | |
Net Weight** | 188 Kg (414 lbs) | 230 Kg (507 lbs) | 288 Kg (634 lbs) |
346 Kg (763 lbs) |
|
Shipping Weight** |
216 Kg (476 lbs) | 285 Kg (628 lbs) | 356 Kg (784 lbs) | 428 Kg (944 lbs) | |
Shipping Dimensions, Maximum (W x D x H)** | 1120 x 820 x 1760 mm (44.0” x 32.3” x 69.3”) | 1450 x 820 x 1760 mm (57.0” x 32.3” x 69.3”) | 1720 x 820 x 1760 mm (68.0” x 32.3” x 69.3”) |
2050 x 820 x 1760 mm (57.0” x 32.3” x 69.3”) |
|
Shipping Volume, Maximum** |
1.62 m3 (57 ft3 ) | 2.09 m3 (74 ft3 ) | 2.48 m3 (87 ft3 ) |
2.96m3 (105 ft3 ) |
AC2-NS Engineering Drawing:
AC2 Cabinets Save Money and Environment:
III. Accessories for microbiological safety cabinet Esco Airstream® Class II Type A2 (AC2-NS)
Accessories for AC2-NS Biological Safety Cabinets |
|||||
Cabinet |
Standard Unit | AC2-3S8-NS 2010946 | AC2-4S8-NS 2010747 | AC2-5S8-NS 2010978 | AC2-6S8-NS 2010963 |
AC2-3S9-NS 2010945 | AC2-4S9-NS 2010752 | AC2-5S9-NS 2010977 |
AC2-6S9-NS 2010925 |
||
Unit with Cable Port |
AC2-3S8-NS-PORT 2011010 | AC2-4S8-NS-PORT 2011011 | AC2-5S8-NS-PORT 2011012 | AC2-6S8-NS-PORT 2011013 | |
AC2-3S9-NS-PORT 2011014 | AC2-4S9-NS-PORT 2011015 | AC2-5S9-NS-PORT 2011016 |
AC2-6S9-NS-PORT 2011017 |
||
Exhaust Ducting |
Anti-blowback Valve 10 inches | ABBV-10P 5170352 | |||
Tri-safe Exhaust Collar with Alarm |
TEM-4 2010606 |
||||
Thimble Exhaust Collar with Alarm |
ECO-A-AC23 5170618 | ECO-A-AC24 5170619 | ECO-A-AC25 5170620 | ECO-A-AC26 5170621 | |
Exhaust Damper |
B2-DAMPER 5170352 |
||||
Hydrogen Peroxide Decontamination Kit |
VHP In Port (Ø 1”) | VHP-IN PORT 5170552 | |||
VHP Out Top Box for Cabinet WITH Exhaust Collar Installed (Ø 1”) | VHP-ECO/OUT 5170554 | VHP-ECO/OUT 5170615 | VHP-ECO/OUT 5170616 |
VHP-ECO/OUT 5170617 |
|
VHP Out Top Box for Cabinet WITHOUT Exhaust Collar Installed (Ø 1”) |
VHP OUT TOP BOX 3ft 5170553 | VHP OUT TOP BOX 4ft 5170612 | VHP OUT TOP BOX 5ft 5170613 |
VHP OUT TOP BOX 6ft 5170614 |
|
Work Zone |
UV Lamp | UV-15A 5170251 | UV-30A-L 5170255 | ||
IV Bar | IV-910 5170499 | IV-1215 5170231 | IV-1520 5170500 |
IV-1825 5170501 |
|
Pre-filter |
PF-41 5090061 | PF-42 5090062 | PF-43 5090063 | PF-44 5090064 | |
Electrical Outlet | Direct Mounted / GFCI |
EO-_ |
|||
Service Fixtures |
EU SF-Gas-20 mm | SF-1G20 5170410 | |||
EU SF-Vacuum-20 mm |
SF-1V20 5170457 |
||||
EU SF-Air-20 mm |
SF-1A20 5170502 | ||||
EU SF-Nitrogen-20 mm |
SF-1N20 5170503 |
||||
EU SF-Water-20 mm |
SF-1W20 5170458 | ||||
EU SF-Universal-22 mm |
SF-2U22 5170504 |
||||
Support Stands, Ships Flat |
Fixed Stand with Leveling Feet, 28” height | SAL-2A0 Gen 2 5130169 | SAL-3A0 Gen 2 5130170 | SAL-4A0 Gen 2 5130134 | SAL-5A0 Gen 2 5130171 |
Fixed Stand with Leveling Feet, 34” height |
SAL-2B0 Gen 2 5130173 | SAL-3B0 Gen 2 5130174 | SAL-4B0 Gen 2 5130175 |
SAL-5B0 Gen 2 5130176 |
|
Fixed Stand with Caster Wheels, 28” height | SPC-2A0 Gen 2 5130161 | SPC-3A0 Gen 2 5130155 | SPC-4A0 Gen 2 5130152 |
SPC-5A0 Gen 2 5130162 |
|
Fixed Stand with Caster Wheels, 34” height |
SPC-2B0 Gen 2 5130164 | SPC-3B0 Gen 2 5130165 | SPC-4B0 Gen 2 5130166 | SPC-5B0 Gen 2 5130167 | |
Telescopic Stand with Leveling Feet, 1” adjustment | STL-2A0 5130092 | STL-3A0 5130050 | STL-4A0 5130051 |
STL-5A0 5130052 |
|
Telescopic Stand with Caster Wheels, 1” adjustment |
STC-2A0 5130135 | STC-3A0 5130055 | STC-4A0 5130056 | STC-5A0 5130057 | |
Motorized Height Stand with Caster Wheels, 39.5” height | SPM-3A2 5130093 | SPM-4A2 5130047 |
SPM-5A2 5130100 |
||
Misc | IQ/OQ Protocol |
9010179 |
IV. Video
***See more videos of microbiological safety cabinets Class I, Class II, Class III at: Gmga Measuring Youtube.No comments yet
SIGN UP FOR A CONSULTATION
Please fill in the form below and send your comments and questions to GMGA, we will respond to your email as soon as possible.