Precision sound calibrator Larson Davis CAL250
Price: Contact
| Model | CAL250 |
| Manufacturer | LARSON DAVIS (a part of the PCB PIEZOTRONICS) |
| Power supply | 6.7 V to 10 V |
| Weight | 156 g |
| Size | 124 x 44.5 mm |
| Condition | Stocking |
| Warranty | 12 months |
IEC 60942 Class 1 Precision sound calibrator Larson Davis CAL250 for 1" microphones at 251.2 Hz Used for Sound Level Meters and Other Sound Measuring Equipment
I. Specifications of Portable sound level calibrator Larson Davis CAL250
1" MICROPHONES - IEC 60942 CLASS 1 CALIBRATOR
Precision sound calibrator Larson Davis CAL250 is a battery operated precision microphone calibrator for 1" microphones used for the calibration of sound level meters and other sound measurement equipment.
The CAL250 delivers a full 114.0 dB level output signal at 251.2 Hz. A 1/2" microphone adaptor is supplied with the CAL250. Adaptors for 1⁄8", ¼", and 3⁄8" microphones are available as optional accessories.
It has been designed for both field and laboratory use and the accuracy has been calibrated to a reference traceable to the National Institute of Standards and Technology (NIST).
The Larson Davis CAL250 features a stable sound pressure independent of the battery condition. In addition, the CAL250 will turn off automatically to preserve battery and guarantee a stable output.
In addition to precision acoustic calibrators, factory calibration services for Larson Davis products are available through the CAL+ program. CAL+ service is provided with all Larson Davis calibrations and includes a complete multi-point factory test, free firmware upgrade to the latest version where applicable, labor warranty extended for one year[1], worn consumables replaced at no charge[2], and more. Contact us for details.
HIGHLIGHTS
- Output level: 114 dB
- Output frequency: 251.2 Hz
- Standard: 1" microphone
- IEC 60942-1:2003 compliant
- Internal battery
- Output level independent of battery condition
- Adaptors: ¹⁄8", 1/4", 3⁄8", 1/2" microphones
- Array adaptor for 8 microphones simultaneously.
TYPICAL APPLICATIONS
Field or laboratory calibration of:
- Sound level meters
- Noise dosimeters
- Noise monitoring stations
- Microphone arrays.
Technical specifications of Precision sound calibrator Larson Davis CAL250
SPECIFICATIONS |
||
|
Acoustic |
||
|
Calibration Sound PressureLevel Initial After 1 Year |
114.0 dB ± 0.1 dB re: 20 μPa ± 0.2 dB re: 20 μPa |
|
|
Equivalent Free-field Level |
0.0 dB for 1/2″ microphones |
|
|
Frequency |
251.2Hz ± 0.7% |
|
|
Harmonic Distortion |
< 2 % |
|
|
Stability After Pressing On |
± 0.1 dB after 2 seconds |
|
|
Minimum Stabilizing Time |
10 seconds after coupling microphone and calibrator |
|
|
Reference Conditions |
101.3 kPa, 23 ºC and 50 % RH |
|
|
Environmental |
||
|
Static Pressure Range |
65 kPa to 108 kPa |
SPL variation < ± 0.25 dB |
|
Temperature Range |
-10 ºC to +50 ºC |
SPL variation < ± 0.4 dB |
|
Humidity Range |
10 % to 90 % RH non-condensing |
SPL variation < ± 0.4 dB |
|
Storage Temperature |
-40 ºC to +60 ºC |
|
|
Storage Humidity |
0 % to 90 % RH (non-condensing) |
|
|
Physical |
||
|
Effective Volume of Calibrator and Microphone |
> 6.1 in3 (100 cm3) |
|
|
Dimensions (L x W x H) |
4.9 x 1.75 in (124 x 44.5 mm) |
|
|
Weight |
5.5 oz (156 g) |
|
|
Power Supply |
||
|
Battery |
9 V NEDA 1604 A or IEC 6LR61 |
|
|
Battery Voltage Operating Range |
6.7 V to 10 V |
|
|
Traceability |
||
|
Traceability |
Traceable to National Institute of Standards and Technology (NIST) |
|
|
Supplied Accessories |
||
|
ADP019 Adaptor for 1/2″ Microphone |
||
|
CCS003 Storage Case |
||
|
9 V Alkaline Battery |
||
|
Users Manual |
||
|
Related Products |
||
|
ADP020 |
Adaptor for 3⁄8″ microphones |
|
|
ADP021 |
Adaptor for 1/4″ microphones |
|
|
ADP023 |
Adaptor for 1⁄8″ microphones |
|
|
ADP031 |
Adaptor for 3⁄8″ microphones |
|
COMPLIANCE |
||
|
Acoustic |
||
|
ANSI S1.40-2006, Class 1 |
||
|
IEC 60942-2003, Class 1 |
||
|
IEC 60942:2017, Class 1 |
||
|
Safety |
||
|
IEC 61010-1:2001 |
||
|
EMC |
||
|
EU directive 2004/108/EC |
||
|
IEC 61326-1:2005 |
||
|
For Use with Microphones of Type |
||
|
IEC 61094-4:1995 |
1″ WS1P, WS1F and WS1D |
|
|
1/2″ WS2P, WS2F and WS2D microphones; with ADP019 adaptor |
||
|
1/4″ WS3P, WS3F and WS3D microphones with ADP021 adaptor |
||
|
According to IEC 61094-1:2000 |
1/2″ LS2P |
|
|
Other Microphones |
3⁄8″ with ADP020 adaptor 1⁄8″ with ADP023 adaptor |
|
|
For Use with Sound Level Meters and Noise Dosimeters |
||
|
ANSI S1.4 Type 1 |
||
|
ANSI S1.25 |
||
|
IEC 61672 Class 1 |
||
|
IEC 61252 |
||
|
[1] Requires regular annual factory calibration. Limited to seven (7) years. [2] Windscreen, O-rings, desiccants, fuses, for example. |
||
No comments yet
SIGN UP FOR A CONSULTATION
Please fill in the form below and send your comments and questions to GMGA, we will respond to your email as soon as possible.

COMMIT
100% new product

COMMIT
100% refund if the product is of bad quality and not as promised

DELIVERY
Nationwide


Review Precision sound calibrator Larson Davis CAL250
There are no reviews yet.